CONSTANT FLOW, CONSTANT PRESSURE SYRINGE DUAL PUMP
HOME » PRODUCTS » CORE ANALYSIS EQUIPMENT » Constant Flow, Constant Pressure Syringe Pump » CONSTANT FLOW, CONSTANT PRESSURE SYRINGE DUAL PUMP

loading

Share to:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
sharethis sharing button

CONSTANT FLOW, CONSTANT PRESSURE SYRINGE DUAL PUMP

RIGCHINA Group Company precision syringe pumps give you flow and pressure control throughout a broad operating range.
From chemicals, to oil/gas, pharmaceutical, and plastics, we continue our tradition of producing quality pumps forever growing markets.
Availability:
Quantity:
  • RCP Seriers

  • RIGCHINA

 Introduction

RIGCHINA Group Company precision syringe pumps give you flow and pressure control throughout a broad operating range. RIGCHINA syringe pumps can be metered with great accuracy and do not exhibit pulsation or flow anomalies, typically associated with other pump types. The proven robust design lowers your down time and service costs to provide years of reliable and worry-free operation

1.1 HANDLE A WIDE VARIETY OF FLUIDS INCLUDING:

  1. 1.    •Aqueous and organic liquids

  2. 2.    •Corrosive solutions

  3. 3.    •Heated fluids

  4. 4.    •Liquefied gases

  5. 5.    •Viscous fluids

  6. 6.    •Slurries and pastes

• Versatile—RIGCHINA pumps can meet more demands than ever before. RIGCHINA syringe pumps are now available in broader pressure ranges.

• Continuous Constant Flow or Pressure— Continuous Constant Flow or Pressure— RIGCHINA pumps operate in either constant fluid delivery or under constant pressure delivery. Operation can be change for different experiments with one key stroke.

• Precision Dosing of Fluids—Flow ranges from 0.00001 to 408 mL/min.

• Broad Operating Temperature Range— Normal operation 5–50 °C ambient. Temperature control jacket will allow you to heat up or cool-down any substance from -30 °C to 100 °C. A high temperature package will allow you to heat your pump up to 200 °C.

• Zero Pulsation—When pulsation cannot be tolerated in experiments, the RIGCHINA pump will deliver. Dual-pump operation offers continuous flow with no pulsation as low as 0.00001 mL/min and up to 20,000 psi of pressure.

 EASY TO USE

Precision control that you demand


“Smart key” programming makes setting up and running your pump system easy and can be learned in just a few minutes. All RIGCHINA pumps, regardless of configuration or operating mode, utilize the same controller, which can be operated up to 15 meters (50 feet) from the pump modules with optional extension cables. Multiple pumps can be controlled with a single program, a configured program, or independently with varied programs. With complete front panel function and front panel accessibility, status, flow rate, and pressure parameters are continuously displayed


From chemicals, to oil/gas, pharmaceutical, and plastics, we continue our tradition of producing quality pumps forever growing markets.


Pump Specifications:

Item

Category

Specification

RCP-100S

RCP-100D

RCP-200S

RCP-200D

RCP-250S

RCP-250D

RCP-500S

RCP-500D

1

Cylinder Capacity

100 ml

Dual Cylinder 100ml*2

200 ml

Dual Cylinder 200ml*2

250 ml

Dual Cylinder 250ml*2

500 ml

Dual Cylinder 500ml*2

2

Uninterruptible work

Refill liquid / 100ml

Uninterruptible work and Independent Operation

Refill liquid

/ 200ml

Uninterruptible work and Independent Operation

Refill liquid

/ 250ml

Uninterruptible work and Independent Operation

Refill liquid / 500ml

Uninterruptible work and Independent Operation

3

FLOW RATE RANGE(ml/min)

0.01 µl/min to   25 ml/min (for any pressure up to 1379 bar)

0.01 µl/min to   50 ml/min (for any pressure up to 1379 bar)

0.01 µl/min to   50 ml/min (for any pressure up to 1379 bar)

0. 1 µl/min to   200 ml/min (for any pressure up to 1379 bar)

4

FLOW RATE   ACCURACY

± 0.3% (maximum   0.25 µl/min seal leakage)

± 0.3% (maximum   0.25 µl/min seal leakage)

± 0.3% (maximum   0.25 µl/min seal leakage)

± 0.3% (maximum   0.25 µl/min seal leakage)

5

FLOW RATE   DISPLAY RESOLUTION

0.01 µl/min (1.0   µl/min in Constant Pressure Mode)

0.01 µl/min (1.0   µl/min in Constant Pressure Mode)

0.01 µl/min (1.0   µl/min in Constant Pressure Mode)

0. 1 µl/min (1.0   µl/min in Constant Pressure Mode)

6

ANALOG OUTPUT   ACCURACY

± 1% of selected   range

± 1% of selected   range

± 1% of selected   range

± 1% of selected   range

7

REFILL TIME

3.5 minutes

7 minutes

8.5 minutes

10.5 minutes

8

REFILL OR   DEPRESSURIZATION RATE

0.01 µl/min to   60 ml/min at any pressure from 0 to 690 bar

0.01 µl/min to   107 ml/min at any pressure from 0 to 518 bar

0.01 µl/min to   135 ml/min at any pressure from 0 to 420bar

0. 1 µl/min to   204 ml/min at any pressure from 0 to 259 bar

9

PRESSURE RANGE (MPa)

0-60 MPa

0-40 MPa

0-100 MPa

0-50MPa

10

PRESSURE   ACCURACY

± 0.5% of full   scale at constant temperature

± 0.5% of full   scale at constant temperature

± 0.5% of full   scale at constant temperature

± 0.5% of full   scale at constant temp.

11

PRESSURE   REPEATABILITY

± 0.5% of full   scale within 48 hours at constant temperature

± 0.5% of full   scale within 48 hours at constant temperature

± 0.5% of full   scale within 48 hours at constant temperature

± 0.5% of full   scale within 48 hours at constant temp.

12

ZERO PRESSURE   DRIFT

± 0.25% of full   scale within 48 hours at constant temperature

± 0.25% of full   scale within 48 hours at constant temperature

± 0.25% of full   scale within 48 hours at constant temperature

± 0.25% of full   scale within 48 hours at constant temp.

13

PRESSURE DISPLAY   RESOLUTION

6.895 kPa

6.895 kPa

6.895 kPa

6.895 kPa

14

DEAD (HEADSPACE)   VOLUME

1.03 ± 0.020 ml

2.00 ± 0.020 ml

2.5 ± 0.020 ml

4.00 ± 0.020 ml

15

Tube Diameter

3 mm

3 mm

3 mm

6 mm

16

Wetted Materials

Hastelloy C-276,

Hastelloy C-276,

Hastelloy C-276,

Hastelloy C-276,

17

AMBIENT TEMP. RANGE

5 to 50°C

5 to 50°C

5 to 50°C

5 to 50°C

18

TEMPERATURE   DRIFT

± 0.0.12% of   full scale/°

± 0.0.12% of   full scale/°

± 0.0.12% of   full scale/°

± 0.0.12% of   full scale/°

19

HUMIDITY

95% maximum

95% maximum

95% maximum

95% maximum


MAXIMUM ALTITUDE

4000 m

4000 m

4000 m

4000 m


Previous: 
Next: