Introduction
RIGCHINA Group Company precision syringe pumps give you flow and pressure control throughout a broad operating range. RIGCHINA syringe pumps can be metered with great accuracy and do not exhibit pulsation or flow anomalies, typically associated with other pump types. The proven robust design lowers your down time and service costs to provide years of reliable and worry-free operation
1.1 HANDLE A WIDE VARIETY OF FLUIDS INCLUDING:
1. •Aqueous and organic liquids
2. •Corrosive solutions
3. •Heated fluids
4. •Liquefied gases
5. •Viscous fluids
6. •Slurries and pastes
• Versatile—RIGCHINA pumps can meet more demands than ever before. RIGCHINA syringe pumps are now available in broader pressure ranges.
• Continuous Constant Flow or Pressure— Continuous Constant Flow or Pressure— RIGCHINA pumps operate in either constant fluid delivery or under constant pressure delivery. Operation can be change for different experiments with one key stroke.
• Precision Dosing of Fluids—Flow ranges from 0.00001 to 408 mL/min.
• Broad Operating Temperature Range— Normal operation 5–50 °C ambient. Temperature control jacket will allow you to heat up or cool-down any substance from -30 °C to 100 °C. A high temperature package will allow you to heat your pump up to 200 °C.
• Zero Pulsation—When pulsation cannot be tolerated in experiments, the RIGCHINA pump will deliver. Dual-pump operation offers continuous flow with no pulsation as low as 0.00001 mL/min and up to 20,000 psi of pressure.
EASY TO USE
Precision control that you demand
“Smart key” programming makes setting up and running your pump system easy and can be learned in just a few minutes. All RIGCHINA pumps, regardless of configuration or operating mode, utilize the same controller, which can be operated up to 15 meters (50 feet) from the pump modules with optional extension cables. Multiple pumps can be controlled with a single program, a configured program, or independently with varied programs. With complete front panel function and front panel accessibility, status, flow rate, and pressure parameters are continuously displayed
From chemicals, to oil/gas, pharmaceutical, and plastics, we continue our tradition of producing quality pumps forever growing markets.
Pump Specifications:
Item | Category | Specification | |||||||
RCP-100S | RCP-100D | RCP-200S | RCP-200D | RCP-250S | RCP-250D | RCP-500S | RCP-500D | ||
1 | Cylinder Capacity | 100 ml | Dual Cylinder 100ml*2 | 200 ml | Dual Cylinder 200ml*2 | 250 ml | Dual Cylinder 250ml*2 | 500 ml | Dual Cylinder 500ml*2 |
2 | Uninterruptible work | Refill liquid / 100ml | Uninterruptible work and Independent Operation | Refill liquid / 200ml | Uninterruptible work and Independent Operation | Refill liquid / 250ml | Uninterruptible work and Independent Operation | Refill liquid / 500ml | Uninterruptible work and Independent Operation |
3 | FLOW RATE RANGE(ml/min) | 0.01 µl/min to 25 ml/min (for any pressure up to 1379 bar) | 0.01 µl/min to 50 ml/min (for any pressure up to 1379 bar) | 0.01 µl/min to 50 ml/min (for any pressure up to 1379 bar) | 0. 1 µl/min to 200 ml/min (for any pressure up to 1379 bar) | ||||
4 | FLOW RATE ACCURACY | ± 0.3% (maximum 0.25 µl/min seal leakage) | ± 0.3% (maximum 0.25 µl/min seal leakage) | ± 0.3% (maximum 0.25 µl/min seal leakage) | ± 0.3% (maximum 0.25 µl/min seal leakage) | ||||
5 | FLOW RATE DISPLAY RESOLUTION | 0.01 µl/min (1.0 µl/min in Constant Pressure Mode) | 0.01 µl/min (1.0 µl/min in Constant Pressure Mode) | 0.01 µl/min (1.0 µl/min in Constant Pressure Mode) | 0. 1 µl/min (1.0 µl/min in Constant Pressure Mode) | ||||
6 | ANALOG OUTPUT ACCURACY | ± 1% of selected range | ± 1% of selected range | ± 1% of selected range | ± 1% of selected range | ||||
7 | REFILL TIME | 3.5 minutes | 7 minutes | 8.5 minutes | 10.5 minutes | ||||
8 | REFILL OR DEPRESSURIZATION RATE | 0.01 µl/min to 60 ml/min at any pressure from 0 to 690 bar | 0.01 µl/min to 107 ml/min at any pressure from 0 to 518 bar | 0.01 µl/min to 135 ml/min at any pressure from 0 to 420bar | 0. 1 µl/min to 204 ml/min at any pressure from 0 to 259 bar | ||||
9 | PRESSURE RANGE (MPa) | 0-60 MPa | 0-40 MPa | 0-100 MPa | 0-50MPa | ||||
10 | PRESSURE ACCURACY | ± 0.5% of full scale at constant temperature | ± 0.5% of full scale at constant temperature | ± 0.5% of full scale at constant temperature | ± 0.5% of full scale at constant temp. | ||||
11 | PRESSURE REPEATABILITY | ± 0.5% of full scale within 48 hours at constant temperature | ± 0.5% of full scale within 48 hours at constant temperature | ± 0.5% of full scale within 48 hours at constant temperature | ± 0.5% of full scale within 48 hours at constant temp. | ||||
12 | ZERO PRESSURE DRIFT | ± 0.25% of full scale within 48 hours at constant temperature | ± 0.25% of full scale within 48 hours at constant temperature | ± 0.25% of full scale within 48 hours at constant temperature | ± 0.25% of full scale within 48 hours at constant temp. | ||||
13 | PRESSURE DISPLAY RESOLUTION | 6.895 kPa | 6.895 kPa | 6.895 kPa | 6.895 kPa | ||||
14 | DEAD (HEADSPACE) VOLUME | 1.03 ± 0.020 ml | 2.00 ± 0.020 ml | 2.5 ± 0.020 ml | 4.00 ± 0.020 ml | ||||
15 | Tube Diameter | 3 mm | 3 mm | 3 mm | 6 mm | ||||
16 | Wetted Materials | Hastelloy C-276, | Hastelloy C-276, | Hastelloy C-276, | Hastelloy C-276, | ||||
17 | AMBIENT TEMP. RANGE | 5 to 50°C | 5 to 50°C | 5 to 50°C | 5 to 50°C | ||||
18 | TEMPERATURE DRIFT | ± 0.0.12% of full scale/° | ± 0.0.12% of full scale/° | ± 0.0.12% of full scale/° | ± 0.0.12% of full scale/° | ||||
19 | HUMIDITY | 95% maximum | 95% maximum | 95% maximum | 95% maximum | ||||
MAXIMUM ALTITUDE | 4000 m | 4000 m | 4000 m | 4000 m |