This browser does not support the video element.
Thickening Time Tests
Thickening time of a slurry is calculated to determine the time required to mix and pump the slurry. Mixing time for the volume is estimated from calculated slurry volume and mixing rate. Mixing rate varies from rig to rig, and is dependent on slurry design. Safety time is allowed to account for any shut down or slow mixing rates that may inadvertently occur.
Atmospheric Consistometer Model RCA-1250
Specifically designed to prepare cement slurries for the testing of various parameters in strict compliance with API Specification 10A/ISO 10426-1:2000 Clause 8 of API Specification 10A outlines the requirements and provides the procedure for conducting the Free-fluid Test (free water) using the Model RCA_1250 Atmospheric Consistometer.
The testing of cement slurries requires the measurement of thickening time, free water content, viscosity, rheological properties, fluid loss and various other properties.
The Model RCA_1250 Atmospheric Consistometer provides a simple method for conditioning the cement slurries in preparation for performing these tests. This instrument is used in laboratories involved in oil well cement research programs, research and testing of cement additives, cement manufacturers quality assurance programs, and in the research for well servicing companies and their field labs.
Features:
l Maximum operating temperature of 100℃ (212°F)
l Unit is operated at atmospheric pressure
l Temperature is maintained via a PID controller
l Process temperature is displayed digitally
l Heat transfer fluid is continuously circulated
l Heater wattage is 1,500
l Slurry container rotational speed is 150 rpm per API specifications
l Dual container design
l Cooling system included
l Stainless steel temperature bath
l Deadweight calibration unit
Atmospheric Consistometer Specifications | |||
Part No. | Model | Description &Technical Parameters | Remark |
1200 | RCA-1200 | Working Temperature: 200° F (93° C) Working Pressure: Atmospheric pressure only Slurry Cup Rotational Speed: 150 rpm Consistency Range: 0 to 100 Bearden Units (Bc) Slurry Container Volume: 28 cu in (470 ml) Electrical Requirements: 115 or 230 VAC 15%; 50/60 Hz 10%,3 KVA Heater: 1,500 watts Dimensions: 25.6 in (65 cm) x 17 in (43 cm) x 18.5 in (47 cm) Weight: 143 lbs (65 kg) | |
1250 | RCA-1250 | Working Temperature: 200° F (93° C) Working Pressure: Atmospheric pressure only Slurry Cup Rotational Speed: 150 rpm Consistency Range: 0 to 100 Bearden Units (Bc) Slurry Container Volume: 28 cu in (470 ml) Electrical Requirements: 115 or 230 VAC 15%; 50/60 Hz 10%,3 KVA Heater: 1,500 watts Dimensions: 25.6 in (65 cm) x 17 in (43 cm) x 18.5 in (47 cm) Weight: 150 lbs (68 kg) |